OpenEx threshold settings lost on mode change
Tech Note: TN0967
Product: OpenEx, SpikePac
Version: All
Date Added: 2016-01-29
Issue
OpenEx/SpikePac threshold settings are lost when changing from preview to record mode.
Workaround
Instead of going directly from Preview to Record, change from Preview to Standby to Record. This gives Controller enough time to register the mode change and save/load the settings properly. This should keep your SpikePac threshold/sorting settings intact.